23 switch pump on/off in a program, Switch pump on/off at an event, Switch on remote control of the suppressor – Metrohm 753 Suppressor Module User Manual
Page 26: Rotate suppressor in a program

3.4 Operation via 732 IC Detector
753 Suppressor Module / 8.753.1003 Instructions for Use
23
Switch pump on/off in a program
In the 732 IC Detector the
create time programs with a maximum of 20 program steps (see
section 4.7.1 of 732/733 Manual). The pump can be switched on
or off at any program step by setting the remote output lead 1 to
1 (on, active, 0 V) or 0 (off, inactive, open) respectively:
Switch on:
>PROGRAM/edit XXX.X min
remote
1
TTTTTTT
Switch off:
>PROGRAM/edit XXX.X min
remote
0
TTTTTTT
Switch pump on/off at an event
In the 732 IC Detector the
program a maximum of 4 different events (see section 4.7.3 of
732/733 Manual). The pump can be switched on or off at any
event by setting the remote output lead 1 to 1 (on, active, 0 V) or
0 (off, inactive, open) respectively:
Switch on:
EVENT YY-MM-DD HH:MM:SS
remote
1
TTTTTTT
Switch off:
EVENT YY-MM-DD HH:MM:SS
remote
0
TTTTTTT
Switch on remote control of the suppressor
ON
OFF
REMOTE
In order to bring the suppressor automatically to the next position
via the 732 IC Detector the remote control of the suppressor must
be switched on with switch 66.
ON
Remote control switched on
OFF
Remote control switched off
Rotate suppressor in a program
In the 732 IC Detector the
create time programs with a maximum of 20 program steps (see
section 4.7.1 of 732/733 Manual ). With each program step the
suppressor can be brought to the next position by setting remote
output lead 2 to 1 (on, active, 0 V). This lead must later be
returned to the setting 0 (off, inactive, open) before the suppres-
sor can be switched to the next position.