beautypg.com

23 switch pump on/off in a program, Switch pump on/off at an event, Switch on remote control of the suppressor – Metrohm 753 Suppressor Module User Manual

Page 26: Rotate suppressor in a program

background image

3.4 Operation via 732 IC Detector

753 Suppressor Module / 8.753.1003 Instructions for Use

23

Switch pump on/off in a program

In the 732 IC Detector the key can be used to

create time programs with a maximum of 20 program steps (see

section 4.7.1 of 732/733 Manual). The pump can be switched on

or off at any program step by setting the remote output lead 1 to

1 (on, active, 0 V) or 0 (off, inactive, open) respectively:

Switch on:

>PROGRAM/edit XXX.X min
remote

1

TTTTTTT

Switch off:

>PROGRAM/edit XXX.X min
remote

0

TTTTTTT

Switch pump on/off at an event

In the 732 IC Detector the key can be used to

program a maximum of 4 different events (see section 4.7.3 of

732/733 Manual). The pump can be switched on or off at any

event by setting the remote output lead 1 to 1 (on, active, 0 V) or

0 (off, inactive, open) respectively:

Switch on:

EVENT YY-MM-DD HH:MM:SS
remote

1

TTTTTTT

Switch off:

EVENT YY-MM-DD HH:MM:SS
remote

0

TTTTTTT

Switch on remote control of the suppressor

ON

OFF

REMOTE

In order to bring the suppressor automatically to the next position

via the 732 IC Detector the remote control of the suppressor must

be switched on with switch 66.

ON

Remote control switched on

OFF

Remote control switched off

Rotate suppressor in a program

In the 732 IC Detector the key can be used to

create time programs with a maximum of 20 program steps (see

section 4.7.1 of 732/733 Manual ). With each program step the

suppressor can be brought to the next position by setting remote

output lead 2 to 1 (on, active, 0 V). This lead must later be

returned to the setting 0 (off, inactive, open) before the suppres-

sor can be switched to the next position.