4.2 Maintenance and servicing

753 Suppressor Module / 8.753.1003 Instructions for Use

27

4.2

Maintenance and servicing

4.2.1

Care

The 753 Suppressor Module requires proper care and attention. Exces-

sive contamination of the instrument could possibly lead to malfunc-

tions and a shorter service life of the inherently rugged mechanical and

electronic parts.
Spilled chemicals and solvents should be wiped up immediately. It is

especially important to protect the plug connections at the rear of the

instrument (particular the mains plug) against contamination.

Although constructional measures have been designed to virtually

eliminate such a situation, should corrosive media penetrate the

interior of the instruments the mains plug of the 753 Suppressor

Module must be immediately disconnected to prevent extensive

damage to the instrument electronics. Inform Metrohm service if your

instrument has been damaged in such a way.

The instrument must not be opened by untrained personnel. Please

comply with the safety notes in section 1.4.1.

4.2.2

Maintenance by Metrohm service

Maintenance of the 753 Suppressor Module is best done as part of an

annual service performed by specialists from the Metrohm company. If

work is frequently performed with caustic and corrosive chemicals, it

may be necessary to shorten the interval between servicing.
The Metrohm service department is always willing to offer expert advice

on the maintenance and servicing of all Metrohm instruments.

4.2.3

Shutdown

If the IC system is shut down for a considerable length of time, the en-

tire IC system (without column and suppressor) must be rinsed free

from salt with methanol/water (1:4) to avoid crystallization of eluent

salts with the corresponding subsequent damage. In order to rinse

them the separating column and suppressor are disconnected and the

injector and detector are directly connected to each other. Rinse with

methanol/water (1:4) until the conductivity drops below 10

µ

S/cm.

All three suppressor units are then rinsed for about 5 min with ultrapure

water.