10 Appendix
870 KF Titrino plus
■■■■■■■■
77
Outputs
Open Collector
tp
t
p
> 200 ms
active = low, inactive = high
I
C
= 20 mA, V
CEO
= 40 V
+5 V: maximum load = 20 mA
10.6.2
Status diagram of the remote interface
[START]
Output 0
Output 1
Output 3
Output 2
Output 4
ready/EOD
condOK
Error
Start
Stop
Input 0
Input 1
EOD
[START]
[STOP]
Titr/Determ
Figure 17
Remote status diagram
EOD = End of Determination
10.7
Remote control via an RS-232 connection
The 870 KF Titrino plus can be remote controlled via an RS-232 connec-
tion. For this, a 6.2148.030 RS-232/USB Box is necessary. Connect the
RS-232/USB Box to the USB interface of the instrument.
The RS-232/USB Box has two connectors for RS-232 instruments. The
RS-232 remote control functions only through the connector RS-232/2.
The connector RS-232/1 is designed for connection with a balance. Con-
nect the RS-232/2 connector of the RS-232/USB Box to a serial interface
(COM interface marked with the symbol IOIOI) of a PC. Use a
6.2134.040 connection cable.