7.1 Karl Fischer Titrations (KFT)
■■■■■■■■■■■■■■■■■■■■■■
62
■■■■■■■■
870 KF Titrino plus
Result unit
Sample size
in …
Factor (FCT)
Divisor (DIV)
mg/mL
mL
1
1
mg/piece
pieces
1
1
Factor (FCT)
Conversion factor, see table above.
Input range
–999999999 - 9999999999
Default value
0.1
Divisor (DIV)
Conversion factor, see table above.
Input range
–999999999 - 9999999999
Default value
1.0
Titer
Titer of the solution used. As soon as a solution has been selected under
Titration parameters, the value and the unit will be read out from the
solution data under System
▶ Solutions and displayed here. If the titer is
changed here manually, then the solution data will be updated.
NOTE
Only the numerical value is taken into account in the calculation for-
mula. The calculation will then provide a correct result only if the unit
of the titer is mg/mL.
Input range
0.00000001 - 9999999999
Default value
1.000
Blank (CV01/2)
This parameter is visible only with the methods KFT Ipol–Blank and KFT
Upol–Blank.
This value is calculated with the blank value determination (methods
Blank Ipol and Blank Upol) and entered here. The value can also be
modified manually.
Input range
–999999999 - 9999999999 mL
Default value
0.0 mL
Decimal places
Number of decimal places used to display the result.