Metrohm 870 KF Titrino plus Manual User Manual
Page 74

7.1 Karl Fischer Titrations (KFT)
■■■■■■■■■■■■■■■■■■■■■■
64
■■■■■■■■
870 KF Titrino plus
Results
The result report contains the calculated results, equivalence points and
endpoints, sample data, etc.
Selection
on | off
Default value
off
Curve
Curve report. The width of the curve is defined in the system settings (see
"Graphics width", page 46).
Selection
on | off
Default value
off
Calculations/Statistics
Output of the calculation formulas for the individual results. The results
are printed with the maximum resolution. This makes recalculation with
an external program possible. The following data is printed out in addition
when statistics is switched on:
■
Result and sample size of the individual determinations
■
Mean value as well as absolute and relative standard deviation
Selection
on | off
Default value
off
Measuring point list
Output of the measuring point list.
Selection
on | off
Default value
off
Parameters
All of the parameters of the current method are printed out with the para-
meter report.
Selection
on | off
Default value
off
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 46).
The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt.