9 Troubleshooting
870 KF Titrino plus
■■■■■■■■
67
9 Troubleshooting
9.1
Karl Fischer titration
Problem
Cause
Remedy
The drift is very high
during conditioning.
The titration cell is leaking.
■
Check seals and septum. If necessary,
replace.
■
Replace molecular sieve.
The drift becomes
greater after each
titration.
The sample releases water
very slowly.
■
Adjust the method.
■
Add solubilizer.
■
Increase the temperature (possibly using a
KF oven).
■
See technical literature.
A side reaction is taking
place.
■
Use special reagents.
■
Adjust the method (increase/decrease the
temperature, external extraction).
■
See technical literature.
The pH value is not in the
optimum range anymore.
Add buffer, see technical literature.
The titration will not
be finished.
The titration cell is leaking.
■
Check seals and septum. If necessary,
replace.
■
Replace molecular sieve.
The minimum increment is
too low.
Define Titration rate = user and increase the
minimum volume increment (Min. incre-
ment) (see Chapter 7.1.3, page 53).
The stop criterion is unsuit-
able.
Adjust the control parameters (see Chapter
7.1.3, page 53):
■
Increase the stop drift.
■
Select a short delay time.
See also: The drift increases
after each titration.