beautypg.com

2 editing the live parameters, Editing the live parameters – Metrohm 870 KF Titrino plus Manual User Manual

Page 38

background image

5.6 Live modifications

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

870 KF Titrino plus

5.6.2

Editing the live parameters

Certain method parameters can be edited while a determination is being
carried out. The only parameters that can be modified are those that can
be selected. Nevertheless, all of the parameters are visible. The modified
parameters are taken into account at once. If you modify, for instance,
the start conditions after the start volume has been dosed, then these
modifications will not be taken into account until the next determination.

Proceed as follows to edit the parameters:

1 Displaying the main dialog

Press [BACK].

The main dialog is displayed. The determination continues to run in
the background.

2 Opening the main menu

Select Menu and press [OK].

3 Editing the method parameters

Select the menu item Parameters and press [OK].

Change the desired parameters accordingly.

4 Displaying the live dialog

Select the menu item Live dialog in the main menu and press
[OK].

or

Press [BACK] in the main dialog.

The live dialog is displayed once again.