7.1 Karl Fischer Titrations (KFT)
■■■■■■■■■■■■■■■■■■■■■■
56
■■■■■■■■
870 KF Titrino plus
Selection
drift | time | rel. drift | off
Default value
drift
drift
The titration is stopped when the stop drift has been reached.
time
The titration is stopped when the endpoint has been exceeded during
a certain period of time (Delay time).
rel. drift
The titration is stopped when the sum of the drift at the start of the
titration and the relative stop drift has been reached.
off
The titration will not be stopped until the stop conditions have been
fulfilled.
Stop drift
This parameter is visible only when Stop criterion = drift.
Titration is stopped when the endpoint and the stop drift have been
reached.
Input range
1 - 999 µL/min
Default value
20 µL/min
Delay time
This parameter is visible only when Stop criterion = time.
When the endpoint is reached, the specified time is allowed to elapse
after the last dosing and the titration is then stopped.
Input range
0 - 999 s
Default value
10 s
Relative stop drift
This parameter is visible only when Stop criterion = rel. drift.
The titration is stopped when the endpoint and the sum of the drift at the
start of the titration and the relative stop drift have been reached.
Input range
1 - 999 µL/min
Default value
10 µL/min
7.1.4
Titration parameters
Menu
▶ Parameters ▶ Titration parameters
Under Titration parameters, the parameters influencing the sequence
of the entire titration are defined.