7.1 Karl Fischer Titrations (KFT)

■■■■■■■■■■■■■■■■■■■■■■

56

■■■■■■■■

870 KF Titrino plus

Selection

drift | time | rel. drift | off

Default value

drift

drift
The titration is stopped when the stop drift has been reached.
time
The titration is stopped when the endpoint has been exceeded during
a certain period of time (Delay time).
rel. drift
The titration is stopped when the sum of the drift at the start of the
titration and the relative stop drift has been reached.
off
The titration will not be stopped until the stop conditions have been
fulfilled.

Stop drift

This parameter is visible only when Stop criterion = drift.

Titration is stopped when the endpoint and the stop drift have been
reached.

Input range

1 - 999 µL/min

Default value

20 µL/min

Delay time

This parameter is visible only when Stop criterion = time.

When the endpoint is reached, the specified time is allowed to elapse
after the last dosing and the titration is then stopped.

Input range

0 - 999 s

Default value

10 s

Relative stop drift

This parameter is visible only when Stop criterion = rel. drift.

The titration is stopped when the endpoint and the sum of the drift at the
start of the titration and the relative stop drift have been reached.

Input range

1 - 999 µL/min

Default value

10 µL/min

7.1.4

Titration parameters

Menu

Parameters Titration parameters

Under Titration parameters, the parameters influencing the sequence
of the entire titration are defined.