10 appendix, 1 exchange unit, 1 maximum dosing and filling rate – Metrohm 870 KF Titrino plus Manual User Manual

Page 80: 2 parameters for the preparing (prep), Maximum dosing and filling rate, Parameters for the preparing (prep)

background image

10.1 Exchange unit

■■■■■■■■■■■■■■■■■■■■■■

70

■■■■■■■■

870 KF Titrino plus

10 Appendix

10.1

Exchange unit

10.1.1

Maximum dosing and filling rate

The maximum dosing rate and maximum filling rate for the exchange unit
depend on the cylinder volume:

Cylinder volume

Maximum rate

1 mL

3.00 mL/min

5 mL

15.00 mL/min

10 mL

30.00 mL/min

20 mL

60.00 mL/min

50 mL

150.00 mL/min

Independent of the cylinder volume, values ranging from 0.01 to 166.00
mL/min can always be entered. When the function is carried out the rate
will be, if necessary, decreased automatically to the highest possible value.

10.1.2

Parameters for the preparing (PREP)

The PREP function (Preparing) is used to rinse the cylinder and tubings of
the exchange unit and fill them air bubble-free. You should carry out this
function before the first determination or once per day.

Preparing is carried out with the following, non-alterable settings:

The entire cylinder volume is dosed twice with the maximum dosing
rate.