7 Parameters
870 KF Titrino plus
■■■■■■■■
53
Pause
Waiting time, e.g. in case of a slow response of the electrode after the
start or reaction time after the dosing of a start volume.
Input range
0 - 999999 s
Default value
0 s
Request sample ID
Selection of the sample identification that is queried in the method run.
Selection
off | ID1 | ID2 | ID1&ID2
Default value
off
Request sample size
If this parameter is switched on, then the value for the sample size will be
queried.
Selection
on | off
Default value
on
For the methods "Blank Ipol" and "Blank Upol" the
default value is off.
Request sample unit
If this parameter is switched on, then the unit for the sample size will be
queried.
Selection
on | off
Default value
off
Hold at request
If this parameter is switched on, then the method run will be paused dur-
ing the query. If the parameter is switched off, the titration will be started
in the background.
Selection
on | off
Default value
on
7.1.3
Control parameters
Menu
▶ Parameters ▶ Control parameters
Under Control parameters, the control parameters for the endpoint are
defined.
Endpoint at
Measured value for the endpoint.