beautypg.com

3 control parameters, Control parameters – Metrohm 870 KF Titrino plus Manual User Manual

Page 63

background image

■■■■■■■■■■■■■■■■■■■■■■

7 Parameters

870 KF Titrino plus

■■■■■■■■

53

Pause

Waiting time, e.g. in case of a slow response of the electrode after the
start or reaction time after the dosing of a start volume.

Input range

0 - 999999 s

Default value

0 s

Request sample ID

Selection of the sample identification that is queried in the method run.

Selection

off | ID1 | ID2 | ID1&ID2

Default value

off

Request sample size

If this parameter is switched on, then the value for the sample size will be
queried.

Selection

on | off

Default value

on
For the methods "Blank Ipol" and "Blank Upol" the
default value is off.

Request sample unit

If this parameter is switched on, then the unit for the sample size will be
queried.

Selection

on | off

Default value

off

Hold at request

If this parameter is switched on, then the method run will be paused dur-
ing the query. If the parameter is switched off, the titration will be started
in the background.

Selection

on | off

Default value

on

7.1.3

Control parameters

Menu

Parameters Control parameters

Under Control parameters, the control parameters for the endpoint are
defined.

Endpoint at

Measured value for the endpoint.