3 file management, File management – Metrohm 870 KF Titrino plus Manual User Manual
Page 54
![background image](/manuals/586441/54/background.png)
6.3 File management
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
870 KF Titrino plus
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.
Date titer det.
Date of the last titer determination.
Monitoring
Activating and deactivating the titer monitoring.
Selection
on | off
Default value
off
Time interval
This parameter is visible only when Monitoring = on.
You will be notified that this time interval (in days) has elapsed when start-
ing a method. You can then select whether or not you would still like to
start the method.
Input range
1 - 999 d
Default value
999 d
6.3
File management
Menu
▶ System ▶ File management
NOTE
This menu item is visible only if a USB flash drive has been connected as
an external storage medium.
Methods can be imported and deleted from a USB flash drive in this dia-
log. Only methods located in the Files directory are displayed in the list
(see "Directory structure on the USB flash drive", page 45).
A backup of the system can be created (all data and settings). Similarly, an
existing backup can be reloaded.
Import
Import the selected method.
Delete
Delete the selected method.
Backup
Create a backup of all data and settings on the USB flash drive.