7.1 Karl Fischer Titrations (KFT)
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
870 KF Titrino plus
Drift value
This parameter is visible only when Drift correction = manual.
Volume drift for manual drift correction.
Input range
0.0 - 99.9 µL/min
Default value
0.0 µL/min
Cond. stop volume
Maximum permissible volume that can be dosed during conditioning.
Conditioning is stopped when the specified volume is dosed. If condition-
ing is continued by pressing [START] once again, then the titrant volume
that has already been dosed will not be taken into account; i.e. the dosing
starts again at zero. The stop volume should be adjusted to the size of the
titration cell in order to prevent any overflow.
Input range
0.00000 - 9999.99 mL
Default value
20.0000 mL
Selection
off
Cond. stop time
Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.
Input range
0 - 999999 s
Selection
off
Default value
off
7.1.2
Start conditions
Menu
▶ Parameters ▶ Start conditions
The parameters that are carried out before the start of titration are
defined under Start conditions.
Start volume
Volume that is dosed prior to the start of the titration.
Input range
0.00000 - 9999.99 mL
Default value
0.00000 mL
Dosing rate
Rate at which the start volume is dosed. The maximum dosing rate
depends on the cylinder volume (see Chapter 10.1.1, page 70).
Input range
0.01 - 166.00 mL/min
Selection
max.
Default value
max.