Metrohm 870 KF Titrino plus Manual User Manual

Page 51

background image

■■■■■■■■■■■■■■■■■■■■■■

6 System settings

870 KF Titrino plus

■■■■■■■■

41

Routine
Limited dialog for routine operations.

Contrast

The contrast of the display can be adjusted with the arrow keys [

] and

[

].

[

]: the contrast will be decreased by one step each time the key is

pressed.

[

]: the contrast will be increased by one step each time the key is

pressed.

Input range

150 - 240

Default value

212

NOTE

Alternatively, the contrast can also be modified in the following man-
ner:

Keep the red [STOP] key pressed down. As soon as the progress bar
appears, also press the arrow key [

] or [] repeatedly.

This method will, however, cause the contrast to be modified by several
steps.

Beep

If this parameter is activated, then a short beep will sound in the following
cases:

When a key is pressed.

At the end of the determination.

When the system remains conditioned without interruption for 10 sec-
onds.

Selection

on | off

Default value

on

PREP warning

If this parameter is activated, then the recommendation will be made to
carry out the function PREP (Preparing):

After the instrument is switched on.

Each time a buret unit is attached.

All tubing and the cylinder are rinsed with this function (see Chapter
10.1.2, page 70)
.