6 calculations - methods "blank ipol/upol, Calculations - methods "blank ipol/upol – Metrohm 870 KF Titrino plus Manual User Manual

Page 69

background image

■■■■■■■■■■■■■■■■■■■■■■

7 Parameters

870 KF Titrino plus

■■■■■■■■

59

Input range

0.00000 - 9999.99 mL

Default value

100.000 mL

Selection

off

Stop time

The titration is stopped when the specified time has elapsed following the
termination of the start conditions.

Input range

0 - 999999 s

Selection

off

Default value

off

Filling rate

Rate at which the dosing cylinder is filled after the titration. The maximum
filling rate depends on the cylinder volume (see Chapter 10.1.1, page
70)
.

Input range

0.01 - 166.00 mL/min

Selection

max.

Default value

max.

7.1.6

Calculations - Methods "Blank Ipol/Upol"

Menu

Parameters Calculation

Calculation formula

EP1*FCT

The calculation formula is predefined and cannot be edited.

The calculated result is saved as a variable Blank CV01 or Blank CV02
for each measuring mode separately (see page 62).

Factor (FCT)

If a larger amount of solvent is used for determining the blank value than
is later used for the sample, then this factor must be used to convert the
endpoint volume accordingly.

Input range

–999999999 - 9999999999

Default value

1.0

Decimal places

Number of decimal places used to display the result.

Input range

0 - 5

Default value

2