6 calculations - methods "blank ipol/upol, Calculations - methods "blank ipol/upol – Metrohm 870 KF Titrino plus Manual User Manual
Page 69
■■■■■■■■■■■■■■■■■■■■■■
7 Parameters
870 KF Titrino plus
■■■■■■■■
59
Input range
0.00000 - 9999.99 mL
Default value
100.000 mL
Selection
off
Stop time
The titration is stopped when the specified time has elapsed following the
termination of the start conditions.
Input range
0 - 999999 s
Selection
off
Default value
off
Filling rate
Rate at which the dosing cylinder is filled after the titration. The maximum
filling rate depends on the cylinder volume (see Chapter 10.1.1, page
70).
Input range
0.01 - 166.00 mL/min
Selection
max.
Default value
max.
7.1.6
Calculations - Methods "Blank Ipol/Upol"
Menu
▶ Parameters ▶ Calculation
Calculation formula
EP1*FCT
The calculation formula is predefined and cannot be edited.
The calculated result is saved as a variable Blank CV01 or Blank CV02
for each measuring mode separately (see page 62).
Factor (FCT)
If a larger amount of solvent is used for determining the blank value than
is later used for the sample, then this factor must be used to convert the
endpoint volume accordingly.
Input range
–999999999 - 9999999999
Default value
1.0
Decimal places
Number of decimal places used to display the result.
Input range
0 - 5
Default value
2