beautypg.com

4 printer, 5 system initialization, Printer – Metrohm 870 KF Titrino plus Manual User Manual

Page 84

background image

10.5 System initialization

■■■■■■■■■■■■■■■■■■■■■■

74

■■■■■■■■

870 KF Titrino plus

Key of the 870 KF Titrino plus

Mouse function

[BACK]

Right mouse button

[

] [] [] []

Mouse movement vertical/hori-
zontal

[

] []

Scroll wheel vertical

10.4.4

Printer

The range of USB printers available is extremely varied and constantly
changing. The following points must be taken into account when select-
ing a printer:

USB interface necessary

Printer language: HP-PCL, Canon BJL Commands, Epson ESC P/2 or
ESC/POS

NOTE

Inexpensive printers are often designed solely for use with a PC and
may not be equipped with one of the printer languages listed above.
Such models are not suitable for this reason.

10.5

System initialization

In very rare instances, a faulty file system (e.g. because of a program
crash) may lead to an impairment of program functioning. The internal file
system must be initialized in such cases.

CAUTION

All user data (methods, solutions, etc.) are deleted if a system initializa-
tion is carried out. Afterwards, the instrument will have the factory set-
tings again.

We recommend creating a backup of the system at regular intervals in
order to avoid data losses.

After a system initialization the program versions and language files do
not have to be reloaded. Only the selection of the dialog language may
have to be reset in the system settings.