beautypg.com

Before using this unit, Warning, Requirements for installation – Yamato Scientific DNF 910 Constant Temperature Ovens User Manual

Page 8: Always ground this unit, Choose a proper place for installation

background image

4

Before Using this unit

Requirements for Installation

WARNING!

1. Always ground this unit

• The DNF400 and DNF600 types use a 100V power source.
• The DNF600 type does not contain a power plug. Please consult your local electrical

contractor for power connecting work.

• The DNF410, DNF610, DNF810 and DNF910 types use a single-phase 200V power source.

Please consult your local electrical contractor for power connecting work.

• Be sure to connect the earth wire (the green cable of power cord) to the grounding conductor

or ground terminal to prevent accidents caused by electric leakage.

• Do not connect the earth wire to gas or water pipes. If not, fire disaster may be caused.
• Do not connect the earth wire to the ground for telephone wire or lightning conductor. If not,

fire disaster or electric shock may be caused.

• Do not use a branching receptacle, which may cause the heat generation.

2. Choose a proper place for installation

• Do not install this unit in a place where:

♦ Rough or dirty surface.

♦ Flammable gas or corrosive gas is generated.

♦ Ambient temperature exceeds 35°C.
♦ Ambient temperature fluctuates violently.

♦ There is direct sunlight.

♦ There is excessive humidity and dust.
♦ There is a constant vibration.

• Install this unit on a stable place

with the space as shown below.
The exhausted opening is provided
on the back surface. Keep away
from it during operation.

DNF400/410
65cm or more
DNF600/610
80cm or more

or more

30cm

30cm or more

30cm

or

more

15cm

or

more

or more

30cm

30cm or more

80cm or more

30cm or more

30cm or more

or more

30cm

70cm or more

DNF400/410/600/610

DNF810

DNF910