Operation Method

Input Program; Insert Step

1

Step insertion procedure

Select the "Insert step" and then press the ENTER key. A step is added there.

Insert step

S03

Set TEMP

►0℃

S03

ENTER

(Selection screen)

(Selection screen for

step inserted)

™

The newly added step is inserted into the step number which is displayed on the selection screen. The

sequence number of each registered step hereafter increases according to the number of steps added.

™

The initial settings below are set in the newly inserted step. The settings of "un-used" steps displayed

at the end of "Select step" are the same as that listed below.

Item Initial

setting

Setting temperature

0℃

Setting time

0 minute

Rotation speed of fan

10

Repeat initiating step

No


Select the time using "▼▲" and press the ENTER key. The setting is decided and the screen

returns to the setting item selection screen.

Press the PROGRAM key to display the next selection screen after the step insertion is