beautypg.com

Operation method, Input program – Yamato Scientific DNF 910 Constant Temperature Ovens User Manual

Page 29

background image

25

Operation Method

Input Program

1

Select program function
• Check that the power in turned on.

PROGRAM

AUTO START

FIXED TEMP

AUTO STOP

PROGRAM

STANDBY

END

HEATER

FAN

FUNCTION

PROGRAM

MENU

ENTER

CANCEL

SUBMENU

POWER

TROUBLE

Step

S01

Start in

30min

DOOR

Press the PROGRAM key. The program menu

starts and step number selection screen is
displayed.
Press the "▼▲". The registered step numbers
and the smallest number of un-used step are
displayed in sequence. Select the step number
among them.

™

The "S01" is displayed when no program is

registered. In this case, the "▼▲" are invalid.

Press the ENTER key. The selected step

number is decided and the setting item selection
screen is displayed. The "Set TEMP" is
displayed first.

Press the CANCEL key to cancel the program

menu.

2

Edit step
• After the step number is decided, the setting item selection screen is displayed. Select the items

on the Sub Display using the PROGRAM key.

No. Item

Sub

Display

Notes

1

Setting

temperature

Set TEMP

100℃

S01

Settable value: 0 to 260℃
(270℃ maximum)
Use with the temperature of 260℃ or less
regardless of the setting temperature range.

2

Setting time

Set time

0min

S01

Settable value: 0 min to 59min, 1h00min to
999h59min, or "End"

3

Rotation speed

of fan

Fan

10

S01

Settable value: 1 to 10, OFF

4

Repeat

initiating step

Rep. start

S01

S01

Input the step number of repeat initiating step
or "No".

5

Repeat count

Rep. count

5

S01

Settable value: 1 to 9999, "Endless"

6

Step insertion

Insert step

S01

Add a new step at the position of step
currently referred to. The sequence number
of each step hereafter increases by one.