beautypg.com

Operation method, Program creation method – Yamato Scientific DNF 910 Constant Temperature Ovens User Manual

Page 41

background image

37

Operation Method

Program Creation Method

1

Edit step

• The unit operates with fixed rate inclination if the "START TEMP" and the "Set Temp" are different.

It enters into "Wait" if the temperature does not reach to the setting temperature within a setting
time.

Set Temp.

TEMP

.

TIME

Set Time

TEMP

.

START

.

(Operation with fixed rate inclination)


• The temperature is kept constant till the end of setting time when the "START TEMP" and the "Set

Temp" are the same. The unit enters into "Wait" and measurement of residual time is suspended
when the temperature in furnace is 6℃ over or 3℃ below to the setting temperature.

Set Temp.

TEMP.

TIME

Set Time

TEMP

.

START

.


• The unit operates with full-power from the "START TEMP" to the " Set Temp". It keeps the "Wait"

state until the temperature in furnace comes over 6℃ or below 3℃ to the setting temperature.

Set Temp.

TEMP.

TIME

Reached time to the set temp.

TEMP

.

START

.

Waiting