background image

38

Operation Method

Program Creation Method

2

Edit step

• Refer to the page 55 for the temperature rise/fall time.

They vary depending on the sample used or condition of exhaust opening (open or close). Check
them by conducting operation.

• The conception of repeat is shown in the figure below. The first operation of repeat interval is not

counted as a repeat count.

Step No. S04

Start step: S02
Repeat times: 5

Step No. S01

No repeat

instruction

Step No. S02

No repeat

instruction

Step No. S03

No repeat

instruction

Step No. S05

No repeat

instruction

5 repeating

of interval


Program creation example;

1

Program creation; Step 1 (S01)

TROUBLE

ENTER

CANCEL

DOOR

Select step

S01

Program OPR

Set TEMP

► 0℃

S01


Set TEMP

► 0min

S01


Press the PROGRAM key. The program menu starts

and step number selection screen is displayed.






Press the ENTER key. The setting temperature

selection screen is displayed.

Press the ENTER key. The setting temperature edition

screen is displayed. Set "100℃" using the "▼▲" and
press the ENTER key to decide it.

Press the PROGRAM key. The setting period selection

screen is displayed.

Press the ENTER key. The setting period edition

screen is displayed. Set "25min" using the "▼▲" and
press the ENTER key to decide it.