beautypg.com

Safety device and error code – Yamato Scientific DNF 910 Constant Temperature Ovens User Manual

Page 63

background image

59

In the Event of Failure…

Safety Device and Error Code


This unit has an automatic diagnosis function built in the controller and safety devices independent of the
controller. The following table shows the purpose, operation and corrective actions against error of
safety devices. If an error occurs, the Main Display indicates the error code and the Sub Display displays
the error details/corrective actions with the buzzer sound. Correct the error according to the instruction.

Safety Device

Purpose

State

Display

Cause/Solution

Earth leakage

breaker

To prevent electric

shock, To protect

over current

• Power shut off
• All indications

goes out

None

™

Check the cause by

contacting to our
service division.

Sensor
trouble

detection

To prevent

overheating due to

sensor failure

• Heater circuit

shut off

• Alarm buzzer

activated

TROUBLE lamp blinks

Sensor Error

Call a service

• Temperature sensor is

broken or disconnected.

™

Check the cause by

contacting to our
service division.

Triac

short-circuit

detection

To prevent

overheating due to

impossibility of

heater control

• Heater circuit

shut off

• Alarm buzzer

activated

TROUBLE lamp blinks

SSR Error

Call a service

• Triac is in short-circuit
™

Check the cause by

contacting to our
service division.

Heater

disconnecting

detection

To notice failure in

temperature

control

• Heater circuit

shut off

• Alarm buzzer

activated

TROUBLE lamp blinks

SSR Error

Call a service

• Heater is disconnected.
™

Check the cause by

contacting to our
service division.

Independent

overheating

prevention

device

To prevent

overheating due to

controller failure

• Heater circuit

shut off

• Alarm buzzer

activated

TROUBLE lamp blinks

Overheat Error

Call a service

• False setting of the

independent overheating
prevention device.

™

Set the device correctly.

• Overheating of sample.
™

Reduce the sample.

• Sensor

of

the

independent overheating
prevention device is
disconnected, or failure
in the independent
overheating prevention
circuit.

™

Check the cause by

contacting to our
service division.