Operation Method

Auto Stop Operation

Start operation

AUTO START

FIXED TEMP

AUTO STOP

PROGRAM

STANDBY

END

HEATER

FAN

FUNCTION

PROGRAM

MENU

ENTER

CANCEL

SUBMENU

POWER

TROUBLE

Set TEMP

100℃↑

Stop in

30min

DOOR

Press the ENTER key to decide the setting and

the auto stop operation starts. The blinking
AUTO STOP lamp lights on and the Sub Display
displays the setting temperature and residual time
to operation stop.

The countdown of timer is suspended when the

temperature in furnace is 3℃ or more lower than
the setting temperature, or 6℃ or more higher
than it. In this case, the Sub Display displays
"Wait" with blinking. The time display on the
right side of "Wait" shows the total waiting time
in operation.

Set TEMP

100℃↑

Wait 1min

(Waiting screen)

4

AUTO START

FIXED TEMP

AUTO STOP

PROGRAM

STANDBY

END

HEATER

FAN

FUNCTION

PROGRAM

MENU

ENTER

CANCEL

SUBMENU

POWER

TROUBLE

Stop time

2002/01/01 12:30

DOOR

The operation stops when the residual time

counts zero. The END lamp blinks and the Sub
Display displays the operation finish time when
the operation stops.

™

The wait function is not activated when the auto stop

operation is carried out with "Clock" mode. The
operation stops at specified time.

The fan stops at the same time the operation

finishes when "OFF" is selected at fan function.

™

The setting of fan function ["ON"/"OFF"] takes

priority, regardless of setting for fan operation in the
standby state.

Press the POWER key to quit operation.