Instrument setup exit screen, Display setpoints and readings program, Running the display setpoints and readings program – RKI Instruments Pioneer 4W User Manual
Page 34: Navigating through the channel setup program

30 • Operation
Pioneer Gas Monitor Operator’s Manual
Instrument setup exit screen
This screen saves or cancels any changes you made in the Instrument Setup program.
•
To save changes
and exit the program, press the ENTER button.
A confirmation message displays, then the normal screen displays. The Pioneer is in
normal operation.
•
To cancel changes
and exit the program, press the ESCAPE button until the following
message displays: “INSTRUMENT SET-UP NOT SAVED. USING PREVIOUS
VALUES.”
The normal screen displays. The Pioneer is in normal operation.
Display Setpoints and Readings Program
This section describes the Display Setpoints and Readings program. You can display the
following in this program. (You can also allow the Pioneer to perform a self check
procedure and reset the minimum and maximum readings in this program.)
•
Software versions
•
Time Weighted Average (TWA) and Short Term Exposure Limit (STEL) readings
(for applicable channels)
•
Minimum and maximum readings (for each channel)
•
Alarm Sets (for each channel)
Navigating through the Display Setpoints and Readings Program
Use the reset switch to begin the program, display the setpoints and readings, and exit the
program.
NOTE:
This is a display only program. You cannot change settings or gas readings in the
Display Setpoints and Readings program.
Running the Display Setpoints and Readings Program
This section describes how to display the various setpoints and gas readings that are
included in this program. This section also describes how to display the time and date and
begin the self check procedure.
NOTE:
Once you display one of the screens, the Pioneer automatically returns to the
normal screen in 1
minute if you do not continue or exit the program.
E
N T E R
=
S t o r e
S e t - U p
E
S C A P E
=
P r e v i o u s