Oxygen Transmitter • 26
Pioneer Operator’s Manual (PRELIMINARY)
4.
Press the ESCAPE button. A confirmation message displays, then the normal screen
appears.
NOTE:
Because you performed calibration at the transmitter, it is not necessary to save
calibration values at the Pioneer. The Pioneer does not save calibration values for
remote transmitters.
The purpose of entering the Calibrate program is to disable the status lights,
buzzer, and relays during calibration and so you can view the current reading(s)
on the Fresh Air screen.
5.
Store the components of the calibration kit in a safe and convenient place.
6.
Verify that the applicable display reading(s) increases and stabilizes at 20.9 %VOL.
Parts List
Table 5 lists replacement parts and accessories for the oxygen channel.
Table 5: Parts List
Part Number
Description
06-1248RK
Sample tubing (1 foot)
18-0300RK
Detector junction box (base only)
18-0301RK-01
Detector junction box (includes base, cover, and gasket)
18-0402RK-01
Amplifier junction box
57-0004RK
Analyzer card
57-1060RK
Amplifier (specify target gas when ordering)
65-0601RK
Oxygen detector
65-2411RK
Oxygen transmitter (includes detector, amplifier, and junction box)
71-0027RK-09
Oxygen Transmitter Manual Insert (this document)
81-0076RK-01
Zero air calibration cylinder (34 liter)
81-0078RK
Calibration cylinder (100% nitrogen, 17 liter)
81-0078RK-01
Calibration cylinder (100% nitrogen; 34 liter)
81-1003RK
Regulator (for 17 and 34 liter calibration cylinders)
81-1109RK
Calibration cup