Parts list, Table 6: parts list – RKI Instruments Pioneer 4W User Manual
Page 127

CO Transmitter • 24
Pioneer Operator’s Manual
3.
Press the ESCAPE button three times.
The Save screen appears.
4.
Press the ESCAPE button. A confirmation message displays, then the normal screen
appears.
NOTE:
Because you performed calibration at the transmitter, it is not necessary to save
calibration values at the Pioneer. The Pioneer does not save calibration values for
remote transmitters.
The purpose of entering the Calibrate program is to disable the status lights,
buzzer, and relays during calibration and so you can view the current reading(s)
on the Fresh Air screen.
5.
Store the components of the calibration kit in a safe and convenient place.
6.
Verify that the applicable display reading(s) decreases and stabilizes at 0 PPM.
Parts List
Table 6 lists replacement parts and accessories for the CO channel.
Table 6: Parts List
Part Number
Description
06-1248RK
Sample tubing (order by the foot)
18-0405RK-01
Junction box (without lid)
33-7101RK
Filter (charcoal)
07-0203RK
Retaining boot (for filter)
57-0004RK
Analyzer card
57-1060RK
Amplifier (specify target gas when ordering)
ES-1531-CO
CO replacement sensor
65-2432RK
CO transmitter (includes detector and amplifier)
71-0027RK-07
CO Transmitter Manual Insert (this document)