Calibration file elements, Calibration file elements 61, Items you may edit – MTS SWIFT 30 Sensor User Manual
Page 61
Select a Zero Method
SWIFT 30 Sensors
Setting up the Transducer Interface
61
Calibration File Elements
The following figure shows some elements of the calibration file:
Typical Calibration File
Items you may edit
•
OutputPolarities—defines the polarities of the six outputs. Change these
only if your application requires different polarities from those on the
transducer label.
•
ZeroAlgorithm—selects a zero algorithm for the application.
•
AngleMode—selects the mode used for determining the encoder sine and
cosine.
•
AngleFixed—used for non-spinning applications.
•
AngleOffset— used for spinning applications. Normally you do not need to
change this value.
•
EncoderSize
1
–defines the size of the encoder. The value will vary
depending on whether the application is using telemetry (0) or spinning slip
ring (1).
[SWIFT]
Name=zero 3 after new tire
SerialNum=1234567
Normalization=0
InputSwitches=255
OutputPolarities=40
ZeroAlgorithm=4
AngleMode=0
AngleOffset=82.0898
AngleFixed=0
EncoderSize=0
ZFX1=0.103786
ZFX2=-0.023199
ZFY1=-0.094017
ZFY2=-0.184371
ZFY3=-0.225885
ZFY4=-0.269841
ZFZ1=-0.101343
ZFZ2=-0.045177
KFX1=0.196337
KFX2=0.195604
KFY1=0.169231
KFY2=0.166789
KFY3=0.167277
KFY4=0.16801
KFZ1=0.195849
KFZ2=0.196581
KMX=3.7558
KMYX=2.39805
KMYZ=2.5641
KMZ=3.75824
KFXFY=0.000611
KFXFZ=-0.00315
KFXMX=0.008864
KFXMY=-0.005348
KFXMZ= 0017998
SWIFT file identifier
Description of the file
Serial number of the
Transducer Interface (TI)
Internal calibration for the TI
Polarity of the six outputs
(default = 40)
Zero Algorithm
Angle Mode
Angle Offset
Fixed Angle
Bridge Zeroes
Do Not Modify
Calibration Gains
Do Not Modify
Setting for all input switches
(default = 255, all closed)
S20-33
Encoder Size
1. Some older versions of TI electronics may have firmware versions that are
not compatible with this parameter. If you are not using telemetry, this will
not affect the function of the SWIFT. To allow proper function, load a
current version of TIXFER or delete this line from the calibration file.