beautypg.com

MTS SWIFT 30 Sensor User Manual

Page 36

background image

SWIFT 30 Sensors

36

Transducer Interface

Hardware Overview

When a system zero is initiated, the associated LED is lit. After successful
completion of the system zero, the LED turns off. If there is a failure during the
system zero, the LED will flash at approximately a 1 Hz rate.

Angle Zero and Bridge Zero Switch Functions

Z

ERO

A

LGORITHM

W

HEN

TO

U

SE

A

NGLE

Z

ERO

B

RIDGE

Z

ERO

0
(Preferred method for
non-spinning
applications)

Use this algorithm for
non-spinning (fixed)

applications.

The Angle Zero switch is
non-functional.

Press the Bridge Zero
switch to measure the

static transducer bridge

offsets.

1

Use this algorithm for

spinning applications.
You will need to
mechanically level the

SWIFT sensor.

Press the Angle Zero

switch to read the current
angle and use the value to

set the angle sum to 0.0°.

Press the Bridge Zero

switch to measure the
static transducer bridge

offsets.

3

Use this algorithm for

spinning applications.
You do not need to

mechanically level the
SWIFT sensor.

Both the Angle Zero and Bridge Zero switches are

functional. This algorithm will perform both angle zero

and bridge zero processes regardless of which button
you press.

4
(Preferred method for
spinning applications)

Use this algorithm for
spinning applications.
You will need to

mechanically level the

SWIFT sensor.

Both the Angle Zero and Bridge Zero switches are
functional. This algorithm will perform both angle zero

and bridge zero processes.