MTS SWIFT 30 Sensor User Manual
Page 36
SWIFT 30 Sensors
36
Transducer Interface
Hardware Overview
When a system zero is initiated, the associated LED is lit. After successful
completion of the system zero, the LED turns off. If there is a failure during the
system zero, the LED will flash at approximately a 1 Hz rate.
Angle Zero and Bridge Zero Switch Functions
Z
ERO
A
LGORITHM
W
HEN
TO
U
SE
A
NGLE
Z
ERO
B
RIDGE
Z
ERO
0
(Preferred method for
non-spinning
applications)
Use this algorithm for
non-spinning (fixed)
applications.
The Angle Zero switch is
non-functional.
Press the Bridge Zero
switch to measure the
static transducer bridge
offsets.
1
Use this algorithm for
spinning applications.
You will need to
mechanically level the
SWIFT sensor.
Press the Angle Zero
switch to read the current
angle and use the value to
set the angle sum to 0.0°.
Press the Bridge Zero
switch to measure the
static transducer bridge
offsets.
3
Use this algorithm for
spinning applications.
You do not need to
mechanically level the
SWIFT sensor.
Both the Angle Zero and Bridge Zero switches are
functional. This algorithm will perform both angle zero
and bridge zero processes regardless of which button
you press.
4
(Preferred method for
spinning applications)
Use this algorithm for
spinning applications.
You will need to
mechanically level the
SWIFT sensor.
Both the Angle Zero and Bridge Zero switches are
functional. This algorithm will perform both angle zero
and bridge zero processes.