Option 5, Option 6 – MTS SWIFT 30 Sensor User Manual
Page 53
TISHUNT - Transducer Interface Shunt
SWIFT 30 Sensors
Software Utilities
53
Option 5
Use this option to command a Shunt Cal in the Transducer Interface. This is the
same as pressing the Shunt button on the front panel of the TI. A period is
displayed on the screen for every change in the shunt state. This gives you a
quick view of the progress.
Option 6
A valid shunt calibration should be performed prior to executing this command.
This option allows an easy means of setting the Shunt Calibration Reference
values after calibration. After this option has executed, the file that was written
can then be downloaded (using option 3). Downloading the file will set the Shunt
Reference values to the last measured values, so that subsequent Shunt
Calibrations should pass when the SWIFT system is setup properly, and fail
when not setup properly.
Note
This menu choice should only be used by qualified service personnel.
TI Shunt Cal state 2
TI Shunt Cal state 5
TI Shunt Cal state 7
TI Shunt Cal state 8
TI Shunt Cal state 11
TI Shunt Cal state 14
TI Shunt Cal state 17
TI Shunt Cal state 20
TI Shunt Cal state 23
TI Shunt Cal state 26
TI Shunt Cal state 29
TI Shunt Cal state 32
TI Shunt Cal state 34
TI Shunt Cal state 35
TI Shunt Cal state 38
TI Shunt Cal state 41
TI Shunt Cal state 43
TI Shunt Cal state 44
TI Shunt Cal state 47
TI Shunt Cal state 0
TI Shunt Cal completed.
Read results by selecting appropriate menu option.