TISHUNT - Transducer Interface Shunt

SWIFT 30 Sensors

Software Utilities

53

Option 5

Use this option to command a Shunt Cal in the Transducer Interface. This is the
same as pressing the Shunt button on the front panel of the TI. A period is
displayed on the screen for every change in the shunt state. This gives you a
quick view of the progress.

Option 6

A valid shunt calibration should be performed prior to executing this command.
This option allows an easy means of setting the Shunt Calibration Reference
values after calibration. After this option has executed, the file that was written
can then be downloaded (using option 3). Downloading the file will set the Shunt
Reference values to the last measured values, so that subsequent Shunt
Calibrations should pass when the SWIFT system is setup properly, and fail
when not setup properly.

Note

This menu choice should only be used by qualified service personnel.

TI Shunt Cal state 2

TI Shunt Cal state 5

TI Shunt Cal state 7

TI Shunt Cal state 8

TI Shunt Cal state 11

TI Shunt Cal state 14

TI Shunt Cal state 17

TI Shunt Cal state 20

TI Shunt Cal state 23

TI Shunt Cal state 26

TI Shunt Cal state 29

TI Shunt Cal state 32

TI Shunt Cal state 34

TI Shunt Cal state 35

TI Shunt Cal state 38

TI Shunt Cal state 41

TI Shunt Cal state 43

TI Shunt Cal state 44

TI Shunt Cal state 47

TI Shunt Cal state 0

TI Shunt Cal completed.

Read results by selecting appropriate menu option.