SWIFT 30 Sensors
42
Transducer Interface
Hardware Overview
TI Jumpers
The transducer interface uses circuit board jumpers to establish certain
parameters and make use of various electronic functions. These jumpers are set at
the factory, and should not be reset. The following table is for reference only.
TI PWB Jumper Locations
J
UMPER
S
ETTING
F
UNCTION
X1
1–2
2–3
Excitation setup for shunts FX1, FY1, FY4, and FZ2.
Sets up – (minus) excitation for shunts FX1, FY1, FY4, and FZ2.
Sets up + (plus) excitation for shunts FX1, FY1, FY4, and FZ2.
X2
1–2
2–3
Excitation setup for shunts FX2, FY2, FY43 and FZ1.
Sets up – (minus) excitation for shunts FX2, FY2, FY43 and FZ1.
Sets up + (plus) excitation for shunts FX2, FY2, FY43 and FZ1.
X4
removed
Not used for TI applications (Standard setting).
X5
installed
removed
Enables the ability to download new code to the block of FLASH
RAM where the boot code is stored.
Disables the ability to download new code to the block of FLASH
RAM where the boot code is stored (Standard setting).
X6
installed
removed
Boots to the FLASH memory down-loader.
Boots to the Transducer Interface software (Standard setting).
X2
X1
X4
X5
X6
S20-13