beautypg.com

11 approvals/approval marks, Approvals/approval marks, 9 technical data – JUMO 14597 safetyM STB/STW Ex - Safety Temperature Limiter and Safety Temperature Monitor Operating Manual User Manual

Page 116

background image

9 Technical data

2013-04-01

110

9.11 Approvals/approval marks

9.12 Important information for the probes in Chapter 9.13 to Chapter 9.15

The following should be noted:

-

There is no safe isolation between the sensor and the fitting. As a result, the sensor connections are to be considered
grounded for the safety assessment.

-

Among other requirements, EN 60079-0 stipulates that for EPL Ga, the mass fraction of aluminum must be less than 10 %
for the production of metallic cases. The probe terminal head used by JUMO contains more than 10 % aluminum. There-
fore the terminal head must be secured with a suitable impact protection for the use of EPL Ga (zone 0). The impact protec-

Approval mark

Testing agen-

cy

Certificates/certification

numbers

Inspection basis

Valid for

DIN

DIN CERTCO

STB/STW 1228

DIN EN 14597

All device versions

SIL2, SIL3

TÜV

Nord

(German Tech-
nical Inspec-
tion Agency)

SAS-0190/2006-3, V2.0

DIN EN 61508

All device versions

PL e

DIN EN ISO 13849

All device versions

GL

Germanischer
Lloyd

36 790-11HH

All device versions

Pressure Equipment
Directive

TÜV Nord
(German Tech-
nical Inspec-
tion Agency)

07 202 4834 Z 5077/3/H

Pressure Equipment Di-
rective
97/23/EC

All device versions

ATEX

TÜV Nord
(German Tech-
nical Inspec-
tion Agency)

TÜV 11 ATEX 556139 X

Directive 94/9/EC

All device versions