beautypg.com

7 test voltages according to en 60730, part 1, 8 electrical safety, 9 environmental influences – JUMO 14597 safetyM STB/STW Ex - Safety Temperature Limiter and Safety Temperature Monitor Operating Manual User Manual

Page 114: Test voltages according to en 60730, part 1, Electrical safety, Environmental influences, 9 technical data

background image

9 Technical data

2013-04-01

108

9.7

Test voltages according to EN 60730, Part 1

9.8

Electrical safety

9.9

Environmental influences

Input and output against voltage supply

- With a voltage supply AC 110 to 240 V +10 %/-15 %

3.7 kV/50 Hz

- With a voltage supply AC/DC 20 to 30 V, 48 to 63 Hz

3.7 kV/50 Hz

Clearances/creepage distances

Mains voltage to electronic com-
ponents and probes

6 mm

/ ≥

8 mm

Mains voltage to relays

6 mm

/ ≥

8 mm

Relays to electronic components
and probes

6 mm

/ ≥

8 mm

Electrical safety

According to DIN EN 14597 (DIN EN 60730-2-9)
Overvoltage category III, pollution degree 2

Protection rating I

With internal isolation from SELV electrical circuits

Ambient temperature range

0 to +55 °C

Storage temperature range

-30 to +70 °C

Temperature influence

≤ ±

0.005 %/K dev. from 23 °C

1

for RTD temperature probes

±0.01 %/K dev. from 23 ×C

1

for thermocouple and current

Resistance to climatic conditions

85 % rel. humidity without condensation
(3K3 with extended temperature range according to DIN EN 60721-3-3)