Caution

Caution: dirt sensitive area
Touching the product or parts thereof with bare
hands increases the desorption rate.
Always wear clean, lint-free gloves and use clean
tools when working in this area.

DANGER

DANGER: cleaning agents
Cleaning agents can be detrimental to health and
environment.
Adhere to the relevant regulations and take the
necessary precautions when handling and
disposing of cleaning agents. Consider possible
reactions with the product materials (

→  13).

Precondition

Gauge deinstalled.

6.2.1

Troubleshooting (measuring chamber)

If the cause of the fault is suspected to be in the measuring
chamber, the following checks can be made with an ohmmeter.

Tools / material required

• Allen wrench AF 2
• Pliers for retaining ring
• Ohmmeter