Gauge identification

3MAx-xxx-0x0N
3MAx-xxx-0x0Q

3MBx-xxx-0x0P



100 k

Ω referenced to supply

common
85 k

Ω referenced to supply

common

Status signal (digital output)

Current rating

100 mA

High voltage is ON

+14.5 … +30 V (dc)
(depending on supply voltage)

High voltage is OFF

0 V (dc)

High voltage cut-in, low active, Pin 7 (digital input)

High voltage ON
High voltage OFF

<2.5 V (dc)
>4.0 V (dc)

High voltage cut-in, high active, Pin 8 (digital input)

High voltage ON
High voltage OFF

>11.0 V (dc)
< 5.0 V (dc)

Supply

DANGER

The gauge may only be connected to power sup-
plies, instruments, or control devices that conform
to the requirements of a grounded protective extra-
low voltage (SELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused.

1)

Supply voltage

at the gauge

2)


+14.5 … +30 V (dc)

Ripple

≤1 V

pp

Power consumption

≤2 W

1)

INFICON controllers fulfill this requirement.

2)

The minimum voltage of the power supply unit must be increased

proportionally to the length of the sensor cable.