tina40e1-b (2011-04)

9

Display panel

Background illumination
Dimensions
Pressure units (pressure p)

LCD matrix, 32×16 pixels

two colors red/green
16.0 mm × 11.2 mm
mbar (default), Torr, Pa
(selecting the pressure unit

→ 33)


DANGER

The gauge may only be connected to power supplies, instruments or
control devices that conform to the requirements of a grounded extra-
low voltage (SELV). The connection to the gauge has to be fused
(INFICON controllers fulfill these requirements).

Operating voltage at the gauge

+24 VDC (+20 … +28 VDC)

1)

ripple max. 2 V

pp

Power consumption

Standard
Degas
Emission start (<200 ms)


≤0.5 A
≤0.9 A
≤1.4 A

Power consumption

BCG450
BCG450-SD
BCG450-SP


≤18 W
≤20 W
≤20 W

Fuse necessary

1.25 AT

BCG450-SD requires an additional, separate power supply for the

DeviceNet interface (

→ 21).

Supply voltage at the DeviceNet con-
nector, (Pin 2 and Pin 3)


+24 VDC (+11 … +25 VDC)

Power consumption

<2 W

The gauge is protected against reversed polarity of the supply voltage.

For reasons of compatibility, the expression "sensor cable" is used for all

BCG450 versions in this document, although the pressure reading of the
gauges with fieldbus interface (BCG450-SD and BCG450-SP) is nor-
mally transmitted via the corresponding bus.

Electrical connector

BCG450
BCG450-SD, -SP

D-Sub,15-pin, male
→ 19

→ 20

Measuring cable

shielded, number of conductors de-
pending on the functions used
(max. 15 conductors plus shielding)

Cable length (supply voltage 24 V

1)

)

Analog and fieldbus operation

RS232C operation


≤35 m, conductor cross-section 0.25 mm²
≤50 m, conductor cross-section 0.34 mm²
≤100 m, conductor cross-section 1.0 mm²

≤30 m

Gauge identification

42 k

Ω resistor between Pin 10 and Pin 5

(sensor cable)

1)

Measured at sensor cable connector (consider the voltage drop as function of

the sensor cable length).

Display (

BCG450)

Power supply

Sensor cable connection