3 status lights – INFICON BCG450 ATM to Ultra-High Vacuum Triple Gauge User Manual
Page 39
tina40e1-b (2011-04)
39
The admissible data rate depends on a number of factors such as system pa-
rameters and cable length
→ [4], [6]). It can be set on the gauge or pro-
grammed via DeviceNet.
By means of the "RATE" switch, the data rate can be set to 125
("1"), 250 ("2") or 500 kBaud ("5").
If the switch is in any of the "P" positions, the data rate is pro-
grammable via DeviceNet (
→ [1]).
Two lights (LEDs) on the gauge inform on the gauge status and the current
DeviceNet status.
Light status Description
Dark No
supply
Flashing
red/green
Selftest
Green Normal
operation
Red
Non recoverable error
Flashing red Recoverable error (e.g.
missing DeviceNet power
supply
Light status Description
Dark
Gauge not online:
− Selftest not yet concluded
− No
supply,
→ "STATUS MOD" light
Flashing
green
Gauge online but no communication:
− Selftest concluded but no communication to other nodes
established
− Gauge not assigned to any master
Green
Gauge online; necessary connections established
Flashing red One or several input/output connections in "time out" status
Red
Communication error. The gauge has detected an error that im-
pedes communication via the network (e.g. two identical node ad-
dresses (MAC IC) or "Bus-off")
The gauge is connected to the DeviceNet system via the 5-pin DeviceNet con-
nector (
→ 21).
4.9.2.3 Data Rate Setting
4.9.3 Status Lights
"STATUS MOD"
(gauge status):
"STATUS NET"
(network status):
Electrical connections