1 setting the switching functions – INFICON BCG450 ATM to Ultra-High Vacuum Triple Gauge User Manual

Page 41

background image

tina40e1-b (2011-04)

41

The gauges BCG450-SD and BCG450-SP have two independent, manually setta-
ble switching functions. Each switching function has a floating normally open relay
contact

1)

. The relay contacts are accessible at the sensor cable connector

(

→ 20).

The threshold values of switching functions A and B can be set within the pressure
range 1×10

-9

mbar … 1000 mbar via potentiometers "SETPOINT A" and

"SETPOINT B".

U

Threshold

= 0.75 × (log p

Setpoint

– c) + 7.75


where constant c is pressure unit dependent (

→ Appendix A).

Time t

Mea

sured

val

ue

Off

Off

On

Hysteresis
10% U

Threshold

Switching function

U

Threshold

Measuring signal

(Pressure p)

(Setpoint A, B)

The hysteresis of the switching functions is 10% of the threshold setting.

1)

Both setpoint relays (SP A/B,

→ 20) can be reprogrammed by the user to

work as atmosphere switching function relays (

→ 30 and [1] or [2])

(SP A/B = default setting).


The threshold values of the two switching functions "SETPOINT A" and
"SETPOINT B" are set locally on the potentiometers of the gauge that are acces-
sible via the openings on one side of the gauge housing.

• Voltmeter
• Ohmmeter or continuity checker
• Screwdriver, max. ø2.5 mm


4.11 Switching

Functions

(BCG450-SD, -SP)

4.11.1

Setting the Switching
Functions

Required tools