4 what to do in case of problems – INFICON BCG450 ATM to Ultra-High Vacuum Triple Gauge User Manual
Page 47
tina40e1-b (2011-04)
47
In the event of a fault or a complete failure of the output signal, the gauge can eas-
ily be checked.
• Voltmeter / ohmmeter
• Allen key, AF 2.5
• Spare sensor (if the sensor is faulty)
The output signal is available at the sensor cable connector (Pin 2 and Pin 12).
In case of an error, it may be helpful to just turn off the mains supply and
turn it on again after 5 s.
Problem Possible
cause
Correction
Output signal
permanently
≈0V
Sensor cable defective or
not correctly connected
Check the sensor cable
No supply voltage
Turn on the power supply
Gauge in an undefined
status
Turn the gauge off and on
again after 5 s (reset)
Output signal +0.1 V
Display: "FAIL EL"
EEPROM error
Turn the gauge off and on
again after 5 s (reset)
Replace electronics
Output signal +0.1 V
Display: "FAIL Cap"
Diaphragm sensor error
(sensor defective)
Replace the sensor
(
→ 49)
Output signal +0.3 V
Display: "FAIL Ion"
Hot cathode error
(sensor defective)
Replace the sensor
(
→ 49)
Output signal +0.5 V
Display: "FAIL Pir"
Pirani error
(sensor defective)
Replace the sensor
(
→ 49)
Electronics unit not
mounted correctly on sen-
sor
Check the connections
(Electronics — sensor)
Display: "no Signal" Internal data connection
not working
Turn the gauge off and on
again after 5 s
Replace the electronics
unit
6.4 What to Do in Case of
Problems
Required tools / material
Troubleshooting (BCG450)