In the event of a fault or a complete failure of the output signal, the gauge can eas-
ily be checked.

• Voltmeter / ohmmeter
• Allen key, AF 2.5
• Spare sensor (if the sensor is faulty)

The output signal is available at the sensor cable connector (Pin 2 and Pin 12).

In case of an error, it may be helpful to just turn off the mains supply and
turn it on again after 5 s.

Problem Possible

cause

Correction

Output signal
permanently

≈0V

Sensor cable defective or
not correctly connected

Check the sensor cable

No supply voltage

Turn on the power supply

Gauge in an undefined
status

Turn the gauge off and on
again after 5 s (reset)

Output signal +0.1 V
Display: "FAIL EL"

EEPROM error

Turn the gauge off and on
again after 5 s (reset)
Replace electronics

Output signal +0.1 V
Display: "FAIL Cap"

Diaphragm sensor error
(sensor defective)

Replace the sensor
(

→ 49)

Output signal +0.3 V
Display: "FAIL Ion"

Hot cathode error
(sensor defective)

Replace the sensor
(

→ 49)

Output signal +0.5 V
Display: "FAIL Pir"

Pirani error
(sensor defective)

Replace the sensor
(

→ 49)

Electronics unit not
mounted correctly on sen-
sor

Check the connections
(Electronics — sensor)

Display: "no Signal" Internal data connection

not working

Turn the gauge off and on
again after 5 s
Replace the electronics
unit