beautypg.com

INFICON BCG450 ATM to Ultra-High Vacuum Triple Gauge User Manual

Page 19

background image

tina40e1-b (2011-04)

19

TxD

RxD

Degas

42 k

Ω

13

14

7

8

2

12

5

15

1.25 AT

24V

Degas

Ident.

RS232

10

-

-

-

+

-

1

4

8

9

1

15

D-Sub,15-pin,
female,
soldering side

Common (power GND 24V supply)

Ground (housing, vacuum connection)

Atmospheric

pressure

reached

Measuring

signal

Electrical connection

Pin 1

Relay "Atmosphere reached", n.o. contact

Pin 2

Measuring signal output

0 … +10.13 V

Pin 4

Relay "Atmosphere reached", com contact

Pin 5

Supply common

0 V

Pin 7

Degas on, active high

0 V/+24 V

Pin 8

Supply

+24 V

Pin 10

Gauge identification

Pin 12

Measuring signal common

Pin 13

RS232, TxD

Pin 14

RS232, RxD

Pin 15

Do not connect

Pins 3, 6, 9 and 11 are not connected internally.


Sensor cable connection
BCG450