Technical Data

89

ninb63e1-f (1410)

Printing of results

The printer port prints the result of every test. In hand probe mode the result printed is
“ACCEPT” or “REJECT” followed by date & time and recipe name (if used) and end
Char New Line (0A, LF).<09> (Char Tab, 09) is used as a separator.

For Example: "TEST_ACCE<09>2013-09-04 13:23:03<09>Factory Default<0A>"

After an active probe test cycle the gas analysis value from the ISH2000 is printed. The
printout can be “2.4E+00A<09>2013-09-04 13:23:03<09>Factory Default<0A>”.

See the ISH2000 manual. If the test cycle is rejected by any other test this will be
printed. Followed with date, time and recipe name. Hardware error prints “ERROR”.
For example: “ERROR<09>2013-09-04<09>Factory Default<0A>”.

On the ILS500 (Service/RS232), you can choose if you want to include time and date in
every result from the ILS500 or not. If it’s on the result will be:

"TEST_ACCE<09>2013-09-04 13:23:03<09>Factory Default<0A>"And if it’s
off:“TEST_ACCE<0A>"

Results from ILS500

Results

Explanation

TEST_ACCE Test

accepted

TEST_REJE

Test rejected

USER_FAIL

User has pressed stop

EVAC_FAIL Evacuation

failed

VDEC_FAIL

Vacuum decay test failed

FILL_FAIL

Tracer gas filling failed

PDEC_FAIL

Pressure decay test failed

BLOC_FAIL

Blockage test failed

REFI_FAIL

Tracer gas refill failed

COMM_FAIL

Communication with ISH2000 failed

TEST_STRT

Test cycle started

TEST_DONE

Test cycle finished

FILL_DONE Filling

completed

CALI_STRT

Calibration started

CALI_FAIL Calibration

failed

CALI_DONE

Calibration successful

RECH_DONE

Recipe change done

RECH_FAIL

Recipe change failed

ERROR

Hardware error on ILS500