INFICON Sensistor ILS500 Leak Detection System User Manual
Page 89
Technical Data
89
ninb63e1-f (1410)
Printing of results
The printer port prints the result of every test. In hand probe mode the result printed is
“ACCEPT” or “REJECT” followed by date & time and recipe name (if used) and end
Char New Line (0A, LF).<09> (Char Tab, 09) is used as a separator.
For Example: "TEST_ACCE<09>2013-09-04 13:23:03<09>Factory Default<0A>"
After an active probe test cycle the gas analysis value from the ISH2000 is printed. The
printout can be “2.4E+00A<09>2013-09-04 13:23:03<09>Factory Default<0A>”.
See the ISH2000 manual. If the test cycle is rejected by any other test this will be
printed. Followed with date, time and recipe name. Hardware error prints “ERROR”.
For example: “ERROR<09>2013-09-04<09>Factory Default<0A>”.
On the ILS500 (Service/RS232), you can choose if you want to include time and date in
every result from the ILS500 or not. If it’s on the result will be:
"TEST_ACCE<09>2013-09-04 13:23:03<09>Factory Default<0A>"And if it’s
off:“TEST_ACCE<0A>"
Results from ILS500
Results
Explanation
TEST_ACCE Test
accepted
TEST_REJE
Test rejected
USER_FAIL
User has pressed stop
EVAC_FAIL Evacuation
failed
VDEC_FAIL
Vacuum decay test failed
FILL_FAIL
Tracer gas filling failed
PDEC_FAIL
Pressure decay test failed
BLOC_FAIL
Blockage test failed
REFI_FAIL
Tracer gas refill failed
COMM_FAIL
Communication with ISH2000 failed
TEST_STRT
Test cycle started
TEST_DONE
Test cycle finished
FILL_DONE Filling
completed
CALI_STRT
Calibration started
CALI_FAIL Calibration
failed
CALI_DONE
Calibration successful
RECH_DONE
Recipe change done
RECH_FAIL
Recipe change failed
ERROR
Hardware error on ILS500