INFICON Sensistor ILS500 Leak Detection System User Manual
Page 59

Calibration
59
ninb63e1-f (1410)
Calibration Settings
Calibrate:
At Startup
After Recipe change
Every
test
Calibration Repeat Pause
50
30
s
Calibration Settings
Cal. Leak in test cycle
Prevent Start
Automatic (Active Probe only)
Max. Attempts
5
Cal. Leak Pressure
Set Cal. Leak Pressure
1.00 bar
Calibration Value
Set to the same as stated on the calibration
certificate issued for the leak or gas.
Calibration Unit
Set to the same as stated on the calibration
certificate issued for the leak or gas.
Leak Gas
Set the gas flow for which the Calibration Leak is
defined.
Notice
For more information, refer to the manuals for Sensistor ISH2000.
Calibrate:
At Startup
Calibration is performed or requested whenever
the power is switched on.
After Recipe Change
Calibration is requested every time another recipe
is selected.
Every “XX” Test
Set number of tests between calibration requests.
Calibration Repeat Pause
Set minimum time between calibrations (recovery
time for sensor).
The unit will signal the interval through a pulsating
sound indication together with the text "Calibrate!".
Notice
The best way to find optimal calibration for an application, is to test
different intervals and learn at what interval calibration is optimal for the
specific case.