1 hardware error messages – INFICON Sensistor ILS500 Leak Detection System User Manual
Page 72

72
Troubleshooting
ninb63e1-f (1410)
11.2.1
Hardware Error Messages
Error Message
Reason for Error
Corrective Action*
Detector Power Off
No power to the Leak
Detector.
Check the power cable to
the detector (internal or
external).
Detector Error + Check
Probe and Cable**
Probe cable disconnected. Connect cable.
Detector Error + Check
Sensor, Voltage Error**
Gas sensor damaged.
Replace probe or sensor.
Detector Error + Error
[Driver name]**
Active Probe error.
Consult Active Probe
Manual.
Hardware Error
Vacuum Sensor Error
No power to vacuum
sensor.
Check cable to sensor.
Sensor not connected to
AD.
Check connection to AD.
Damaged vacuum sensor. Send in for repair.
Hardware Error
Pressure Sensor Error
No power to pressure
sensor.
Check cable to sensor.
Sensor not connected to
AD.
Check connection to AD.
Damaged pressure sensor. Replace sensor.
Analog Inputs Power Off
No power to AD module.
Check power cable on left
side of AD module.
Emergency Stop Activated
Emergency Stop not
Reset.
Pull out emergency Stop
Button to reset.
ISH2000 Comm. Error
ISH2000 printer mode
turned off manually.
Restart system.
ISH2000 serial cable
disconnected.
Connect cable
(internal or external).
*
Contact your supplier if the suggested action does not clear the error.
** Error message on ISH2000 display.