beautypg.com

Dakota Ultrasonics MVX User Manual

Page 45

background image

MVX High Performance Thickness Gauge

Note: Be sure that a probe zero has been done prior to performing this

calibration procedure.

1) Using a set of calipers, physically measure two known thickness locations

on the sample of material. Note: The two locations should represent the

desired thickness range being measured. If the range is from .250” - 1.000”,

then the two known locations should represent this range, or be slightly

outside or inside of this range. A typical example would be to find two

locations at .236” and the other at 1.047”. These thickness points will satisfy

the required thickness range noted above.

2) Apply a drop of couplant on the transducer. Place the transducer in steady

contact on thickest known location first. Be sure that the reading is stable

and the repeatability indicator, located in the top left corner of the display, is

fully lit and stable. Press the MENU key once to activate the menu items tab.

Press the MENU key multiple times to tab right, and the ESC key multiple

times to tab left, until the CAL menu is highlighted and displaying the

submenu items.