Dakota Ultrasonics MVX User Manual
Page 16
Dakota Ultrasonics
12
B-SCAN
DIGITS
In the upper left corner of each of the mode photos above, is the repeatability
indicator. The repeatability indicator is represented by six vertical bars and
represents how repeatable the measurements are. In regular measurement mode,
the MVX makes 4 measurements a second. In scan mode, the MVX makes 32
measurements a second. When the MVX is idle, only the left vertical bar and the
underline will be displayed. However, when the MVX is making a measurement, five
or six of the bars should be displayed on the repeatability indicator. If fewer than five
bars are showing, the MVX is having difficulty achieving a stable measurement and
the thickness value displayed is potentially be unstable. Another thing to note is the
vertical broken line in the RF and RECTIFIED photos above. This line represents the
actual point of detection. The MVX uses a zero crossing or flank detection method,
and presents the actual point of measurement on the display.