beautypg.com

Dakota Ultrasonics MVX User Manual

Page 14

background image

Dakota Ultrasonics

10

DIGITS: Displays the digital thickness value using a larger font size. This view is
useful when the MVX is being used as a basic thickness gauge.

Once the view has been selected according to the application requirements, the
delay and width of the screen will potentially need to be adjusted, if the view has
been set to RF, RECT, or BSCAN. Use the following steps to adjust these settings
directly from the measurement screen as follows:

Adjusting the Delay & Width

1)

Press the MEAS key once to activate the measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times to

move left, until the either the DELAY or WIDTH cell is highlighted.

2)

Use the UP, DOWN, LEFT, or RIGHT arrow keys to scroll the DELAY and

WIDTH values.

3)

Repeat steps 1 & 2 until the range is correctly being displayed.

Alternatively, the DELAY and WIDTH values can be changed using the Digit

Edit Box as follows:

1)

Press the MEAS key once to activate measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times to

move left, until the either the DELAY or WIDTH cell is highlighted.