values for water at various temperatures. Once the new value has
been entered on Line 4, press ‘ENTER’.

Pressing the

key will next display ‘→Calibration’ on Line 2.

Press the ‘ENTER’ key and ‘

→C0’ will appear on Line 3 and a

numerical value on Line 4. Press the ‘ENTER’ key and use the
arrow keys

to enter the calibration factors that will accompany the

sensor. Once the ‘C0’ value is entered press the ‘ENTER’ key to
store that value to memory. Line 3 will show ‘

→C0’; press the

key to move to ‘

→C1’, press ‘ENTER’, and use the arrow keys to

enter ‘C1’ values. Once the values are entered, press ‘ENTER
and press the

key to move to ‘→C2’ and enter those values and

press ‘ENTER’ and the ‘BACK’ key.

The display will show ‘

→Calibration’ on Line 2. Press the key to

scroll to ‘

→Flow Direction’ on Line 2. If it is necessary to change

the flow direction (if the sensor was installed with the Flow
Direction arrow on the band opposite the actual flow direction
within the pipe or the process flow has changed direction) press
the ‘ENTER’ key and toggle the

or to change flow direction.

Once changed press the ‘ENTER’ key.

Press the

key to scroll to ‘→Set Date/Time’ on Line 2. Press the

ENTER’ key and the current saved date and time will be displayed
on Line 4 of the display. Use the arrow keys (

and to scroll

through the values and

and to change character) to enter the

date and time. Note: Time is in 24 hour format. Once the date
and time have been set press the ‘ENTER’ key followed by the
BACK’ key.

At this point ‘Basic Config’ is displayed on Line 1 of the display.
The user can re-enter this menu if desired by pressing the
ENTER’ key or by pressing the

or arrows to move to other

Level 1 menus (see Table 14).

Entry of inputs to the other Level 1 menus is by the same process as
used in ‘Basic Config’.

The following table provides a template for recording the transmitter