CiDRA SONARtrac HD VF-100 User Manual
Page 93

Copyright © 2006 CiDRA Corporation
Page 11-2
20638-02 Rev 01
•
‘
→Sensor Serial #’ will appear on Line 2 of the display. Press the
‘ENTER’ key and the current sensor serial number entered in the
transmitter will be displayed on Line 4 of the display. The user can
then use the arrow keys (
↑ and ↓ to scroll through the values and
← and → to change position) to enter the sensor serial number.
Once all digits are entered, press the ‘ENTER’ key to save to
memory. ‘
→Sensor Serial #’ will again be displayed on Line 2.
•
Next, press the
↓ key to scroll to the ‘Pipe Size’ menu on Line 2.
Press the ‘ENTER’ key to access the options available under that
menu. Note: it is necessary to access only one of the following
options.
•
The first choice on the ‘Pipe Size’ menu is ‘
→ID / Wall’ shown on
Line 3 of the display. If this value is known it can be entered here.
Press the ‘ENTER’ key and the current saved inner diameter and
wall thickness will be displayed on Line 4 of the display. Use the
arrow keys (
↑ and ↓ to scroll through the values and ← and → to
change character) to enter the pipe inner diameter. Units can be
inches or millimeters. Once all digits are entered, press the
‘ENTER’ key to save the value to memory.
•
The second choice on the ‘Pipe Size’ menu is ‘Size/Sched’ (pipe
Size / Schedule). If this value is known press the ‘ENTER’ key.
Use the arrow keys to enter the values and then press the
‘ENTER’ key to save to memory.
•
The third choice on the ‘Pipe Size’ menu is ‘OD / Wall’ displayed
on Line 3. Use the arrow keys (
↑ and ↓ to scroll through the values
and
← and → to change character) to enter the pipe inner
diameter. Units can be inches or millimeters. Once all digits are
entered, press the ‘ENTER’ key to save the value to memory.
•
Following the entering of ‘Pipe Size’ press the ‘BACK’ key and
‘
→Pipe Size’ will be displayed on Line 2. At this point the user can
either press the ‘ENTER’ key to re-enter the ‘Pipe Size’ menu or
press the
↓ key to move to the next menu item.
•
Pressing the
↓ key will next display ‘→Fluid Properties’ on Line 2.
Press the ‘ENTER’ key and ‘
→Specific Gravity’ will appear on
Line 3 and the current value on Line 4. To change the Line 4 value
press ‘ENTER’ and use the arrow keys to enter the new value.
Water at 25 ºC (0.997) is the default. Appendix E lists values for
water at various temperatures. Once the new value has been
entered on Line 4, press ‘ENTER’.
•
If no change is made (or after a change in Specific Gravity has
been made) pressing the
↓ key will next display ‘→VISCOSITY (Pa
s)
’ on Line 3 and the current value on Line 4. To change the Line
4 value press ‘ENTER’ and use the arrow keys to enter the new
value. Water at 25 ºC (8.9008 e
-04
) is the default. Appendix E lists