beautypg.com

CiDRA SONARtrac HD VF-100 User Manual

Page 93

background image

Copyright © 2006 CiDRA Corporation

Page 11-2

20638-02 Rev 01

→Sensor Serial #’ will appear on Line 2 of the display. Press the

ENTER’ key and the current sensor serial number entered in the
transmitter will be displayed on Line 4 of the display. The user can
then use the arrow keys (

and to scroll through the values and

and to change position) to enter the sensor serial number.
Once all digits are entered, press the ‘ENTER’ key to save to
memory. ‘

→Sensor Serial #’ will again be displayed on Line 2.

Next, press the

key to scroll to the ‘Pipe Size’ menu on Line 2.

Press the ‘ENTER’ key to access the options available under that
menu. Note: it is necessary to access only one of the following
options.

The first choice on the ‘Pipe Size’ menu is ‘

→ID / Wall’ shown on

Line 3 of the display. If this value is known it can be entered here.
Press the ‘ENTER’ key and the current saved inner diameter and
wall thickness will be displayed on Line 4 of the display. Use the
arrow keys (

and to scroll through the values and and to

change character) to enter the pipe inner diameter. Units can be
inches or millimeters. Once all digits are entered, press the
ENTER’ key to save the value to memory.

The second choice on the ‘Pipe Size’ menu is ‘Size/Sched’ (pipe
Size / Schedule). If this value is known press the ‘ENTER’ key.
Use the arrow keys to enter the values and then press the
ENTER’ key to save to memory.

The third choice on the ‘Pipe Size’ menu is ‘OD / Wall’ displayed
on Line 3. Use the arrow keys (

and to scroll through the values

and

and to change character) to enter the pipe inner

diameter. Units can be inches or millimeters. Once all digits are
entered, press the ‘ENTER’ key to save the value to memory.

Following the entering of ‘Pipe Size’ press the ‘BACK’ key and

→Pipe Size’ will be displayed on Line 2. At this point the user can

either press the ‘ENTER’ key to re-enter the ‘Pipe Size’ menu or
press the

key to move to the next menu item.

Pressing the

key will next display ‘→Fluid Properties’ on Line 2.

Press the ‘ENTER’ key and ‘

→Specific Gravity’ will appear on

Line 3 and the current value on Line 4. To change the Line 4 value
press ‘ENTER’ and use the arrow keys to enter the new value.
Water at 25 ºC (0.997) is the default. Appendix E lists values for
water at various temperatures. Once the new value has been
entered on Line 4, press ‘ENTER’.

If no change is made (or after a change in Specific Gravity has
been made) pressing the

key will next display ‘→VISCOSITY (Pa

s)

’ on Line 3 and the current value on Line 4. To change the Line

4 value press ‘ENTER’ and use the arrow keys to enter the new
value. Water at 25 ºC (8.9008 e

-04

) is the default. Appendix E lists