HD VF-100 PERFORMANCE SPECIFICATIONS
A2.1 Turn-on
Time
30 minutes to rated accuracy from power up
25 seconds from power interruption
A2.2 Start-up
Time
25 seconds from zero flow
A2.3
Low / High Flow Cutoff
Adjustable between 3 and 30 ft/sec. At process flow rates below and
above these values, the output will register a ‘
indication.
A2.4 Flow
Rate
Range
SONARtrac
TM
is capable of measuring signals from process fluids
traveling at velocities between 3 ft/s and 30 ft/s (1.0 m/s to 10 m/s).
This capability is for both forward or reverse flow directions.
A2.5 Accuracy
Flow rate accuracy is +/-1.0% of flow rate from 3 to 30 ft/sec (1 to
10m/s).
A2.6 Repeatability
+/- 0.3 % of reading
A2.7 Update
Rate
2 seconds