20638-02 Rev 01
•
Totalizer
Selections under this menu are used to configure the totalizer
functions.
-
‘Units’ submenu is used to select the units of the display. Currently
these are ‘gal’ (gallons), ‘m
3’
(cubic meters), ‘VF Vol Units’
(volumetric flow display units), ‘l’ (liters) ‘ft
3’
(cubic feet).
-
‘Lowcut Enable’ turns the lowcut function on or off.
-
‘Lowcut’ sets the flow values that will no longer be used for
totalization.
-
‘Multiplier’ selects the multiplier applied to the displayed totalized
units. Selections of ‘M’ (x 1,000,000), ‘k’ (x 1,000), ‘1’ (x 1) are
available. Note: This selection does not affect the pulse output when
‘Totalizer’ is selected. Set the pulse output multiplier in the ‘Pulse’
setup menu.
-
‘Reset’ sets the totalizer display to zero.
•
Wr Protect Mode
When this is ‘Enabled’ no other parameters can be changed. The
user must ‘Disable’ this option before making any menu changes.
Default is ‘Disable’.
10.1.5 Communications
Menu
•
Ethernet
Option is used to view and set the IP address and Subnet Mask of the
transmitter.
•
Serial Options
Used for setting up the ‘Baud Rate’ and the serial communications to
‘RS232’ or ‘RS485’.
•
HART
Settings are used for configuring the protocol when using HART.
Default is for single device addressing.
•
Reset Comms
Used to re-initialize the communications ports without loosing data
history (as would happed by re-initializing the transmitter). This can
be used to restart acquiring data if, for example, the ethernet cable
was removed from the ethernet port while the StBasic program was
still running.