7 info menu, 7 info menu -14 – CiDRA SONARtrac HD VF-100 User Manual
Page 89
Copyright © 2006 CiDRA Corporation
Page 10-14
20638-02 Rev 01
based gain control functions are provided through three sub-menus –
‘AUTOSET GAIN’, ‘CHECK/SET GAIN’, and ‘TEST GAIN’. Gain
adjustments should be made to the system when the process is
operating “normally”.
-
‘AUTOSET GAIN’ – performs an automatic test and adjustment of the
pre-amplifier gain setting. It automatically cycles through the pre-
determined gain settings in order to find the optimal gain setting based
on the flow conditions at that time. AUTOSET GAIN should be run
while the process is operating at normal conditions to avoid making an
incorrect gain setting.
-
‘CHECK/SET GAIN’ – enables the user to check the current gain
setting in the pre-amplifier and to manually set it to one of four user
settings 1, 4.65, 21.55, and 98.65.
-
‘TEST GAIN’ – performs a test of the system electronics to determine
if the gain setting is optimal.
•
Self Test
‘PASS/FAIL’ test performed on the system RAM and DPRAM (internal
memory).
•
Keyboard Test
Tests the function of each of the keyboard keys. The alarm LED will
light for each key pressed to indicate the test is in progress.
•
Clear History
Will delete transmitter stored Data History and re-start saving of
transmitter data.
•
Param Edit
This is a diagnostic feature rarely used and should only be used under
the direction of CiDRA personnel. Contact CiDRA Technical Support.
•
Monitor
Displays a set of various ‘System’ or ‘Sensor’ parameters, updated at
the display update rate. Used by Factory Technical Support
personnel.
10.1.7 Info
Menu
The ‘Info’ menu allows the user to get detailed information on the flow
monitoring system. The
↑ and ↓ keys scroll through the pages in the
Info menu.
•
Revisions
Provides a multi-page list of installed system revisions, hardware /
software serial / model numbers and part numbers.