beautypg.com

2 armored cable installation – CiDRA SONARtrac HD VF-100 User Manual

Page 48

background image

Copyright © 2006 CiDRA Corporation

Page 8-8

20638-02 Rev 01

Table 6

Sensor to Transmitter Cable Terminal Connections

8.4.2.3.2

Armored cable installation

For those installations that will use armored sensor to transmitter
cable, install the P/N 20448-01 transmitter housing stiffener plate
(shipped with the armored cable) in the transmitter housing. Ensure
the plate is installed with the bent edge up and located in the front
portion of the transmitter housing. The stiffener plate is held in place
by the cable gland fittings.

Figure 39

Armored Cable Stiffener Plate Installation

Wire Pair # Transmitter

Terminal #

Function

SENSOR CONNECTOR
PIN NUMBER

1 Wht/Blk

1

Sensor #1 Input

1 - Wht

13 - Blk

2 Wht/Blk

2

Sensor #2 Input

2 - Wht

14 - Blk

3 Wht/Blk

3

Sensor #3 Input

3 - Wht

15 - Blk

4 Wht/Blk

4

Sensor #4 Input

4 - Wht

16 - Blk

5 Wht/Blk

5

Sensor #5 Input

5 - Wht

17 - Blk

6 Wht/Blk

6

Sensor #6 Input

6 - Wht

18 - Blk

7 Wht/Blk

7

Sensor #7 Input

7 - Wht

19 - Blk

8 Wht/Blk

8

Sensor #8 Input

8 - Wht

20 - Blk

9 Wht/Blk

9

Spare – unused

--- ---

10 Wht/Blk

10

Wht – RS 485 Hi / Blk – RS485 Low

12 - Wht

24 - Blk

11 Wht/Blk

11

Wht – ‘-12 V’ / Blk – Gnd

9 - Wht

21 - Blk

12 Wht/Blk

12

Wht – ‘+12 V’ / Blk – Gnd

10 - Wht

22 - Blk

Transmitter box

Armored cable
stiffener plate