2 armored cable installation – CiDRA SONARtrac HD VF-100 User Manual
Page 48
Copyright © 2006 CiDRA Corporation
Page 8-8
20638-02 Rev 01
Table 6
Sensor to Transmitter Cable Terminal Connections
8.4.2.3.2
Armored cable installation
For those installations that will use armored sensor to transmitter
cable, install the P/N 20448-01 transmitter housing stiffener plate
(shipped with the armored cable) in the transmitter housing. Ensure
the plate is installed with the bent edge up and located in the front
portion of the transmitter housing. The stiffener plate is held in place
by the cable gland fittings.
Figure 39
Armored Cable Stiffener Plate Installation
Wire Pair # Transmitter
Terminal #
Function
SENSOR CONNECTOR
PIN NUMBER
1 Wht/Blk
1
Sensor #1 Input
1 - Wht
13 - Blk
2 Wht/Blk
2
Sensor #2 Input
2 - Wht
14 - Blk
3 Wht/Blk
3
Sensor #3 Input
3 - Wht
15 - Blk
4 Wht/Blk
4
Sensor #4 Input
4 - Wht
16 - Blk
5 Wht/Blk
5
Sensor #5 Input
5 - Wht
17 - Blk
6 Wht/Blk
6
Sensor #6 Input
6 - Wht
18 - Blk
7 Wht/Blk
7
Sensor #7 Input
7 - Wht
19 - Blk
8 Wht/Blk
8
Sensor #8 Input
8 - Wht
20 - Blk
9 Wht/Blk
9
Spare – unused
--- ---
10 Wht/Blk
10
Wht – RS 485 Hi / Blk – RS485 Low
12 - Wht
24 - Blk
11 Wht/Blk
11
Wht – ‘-12 V’ / Blk – Gnd
9 - Wht
21 - Blk
12 Wht/Blk
12
Wht – ‘+12 V’ / Blk – Gnd
10 - Wht
22 - Blk
Transmitter box
Armored cable
stiffener plate